Six Degree-of-Freedom Image Guidance for Frameless Intracranial Stereotactic RadioSurgery with kilo-voltage Cone- Beam CT

Abstract

Zheng Chang, Zhiheng Wang, Jinli Ma, Q. Jackie Wu, Scott Senick, Ryan McMahon, John P. Kirkpatrick and Fang-Fang Yin

Abstract Purpose: To investigate localization accuracy for frameless intracranial stereotactic-radiosurgery using 2D orthogonal planar imaging and 3D cone-beam CT (CBCT) in 6-degree-of-freedom (6-DOF).

Methods and Materials: In a phantom study, a target ball phantom was used to perform a Winston-Lutz test and to verify coincidence of imaging isocenter and radiation isocenter. A head phantom was placed with pre-defined positions and imaged with CBCT to test imaging accuracy. In a patient study, one hundred patients were included. Patients were initially positioned with a thermoplastic frameless mask system and then aligned with orthogonal planar imaging and CBCT. The setup discrepancies were quantitatively analyzed.

Results: Phantom experiments showed discrepancies in root-mean-square were 1.6mm translationally and 0.5° rotationally between CBCT 6-DOF image guidance and the known displacements after deviations from the radiation isocenter are considered. In the patient study, setup displacements between orthogonal planar imaging and CBCT 6-DOF image guidance were 3.2mm translationally and 0.9° rotationally. The positioning of twelve patients was corrected in 6-DOF using CBCT and a robotic couch to reduce translational and rotational discrepancies of 1.4mm and 1.3°, as compared with standard CBCT translational correction. Conclusion: CBCT 6-DOF image guidance offers an explicit view to verify patient positioning in translations and rotations.

अस्वीकृति: इस सारांश का अनुवाद कृत्रिम बुद्धिमत्ता उपकरणों का उपयोग करके किया गया है और इसे अभी तक समीक्षा या सत्यापित नहीं किया गया है।

इस लेख का हिस्सा

जर्नल हाइलाइट्स

में अनुक्रमित

arrow_upward arrow_upward